NEW SEARCH NEW SEARCH, STATE - ACT
First Prev Next Last
DISCLAIMER:
Whilst all care is taken to provide accurate information with respect to the item described, the Independent Living Centre (ILC) is not involved in product design or manufacture, and therefore not in a position to guarantee the accuracy of the information provided. Selection of equipment, which is both suitable and appropriate for individual needs remains the responsibility of the person(s) considering requisition, and no responsibility is taken by the ILC for any loss or injury caused through use of the equipment or alleged to have arisen through reliance upon information provided. As information is subject to change any enquiries should be directed to the manufacturer.
Item Details
Autoexcel Micro Air Loss System - Mattress Overlay
ILC Reference NO 21:14:024
Item sourced from ILC ACT database
Short Description
An alternating air mattress overlay that combines micro air loss with pressure relieving action. It is a mattress overlay which can be used in hospital and home environments, especially where excessive moisture build up is a problem and maceration of skin tissue must be reduced.
Designed for use by people who are at high risk of developing pressure ulcers and those with existing tissue damage up to and including Grade 3 Pressure Ulcers. It consists of a portable pump and mattress pad. Available in one size only.
Price Guide
Refer to supplier details for pricing.
More Details
Mattress
The mattress features individual cylindrical air chambers that span the full width of the mattress
The chambers press stud into place and are separate from the base of the overlay and each other so that they can be removed for cleaning and replaced if punctured
CPR release deflates mattress within 15 seconds
Pump
Portable slimline pump with audio/visual alarms,
Quiet and vibration free
Supplies air to each alternate chamber resulting in a ten minute cycle of alternative inflation and deflation
Micro Air Loss
The person's comfort and skin integrity can be severely compromised if excessive moisture is allowed to build up between the support surface and the user.
Micro air loss creates an osmotic gradient moving vapour away from the skin and keeping the person dry and comfortable
The gentle circulation helps in the reduction of humidity
Three air chambers at the head end remain at a constant pressure to minimise the sensation of movement for the user
Automatic Alternating Action
Gentle inflation and deflation for the support cells, allows oxygenation and nutrition of tissue to occur, preventing breakdown and enhancing healing
An Auto-Matt sensor pad automatically adjusts according to the user's movement and weight distribution thus providing optimum support for the person
Cover
Dartex cover which is vapour permeable, water resistant, bacteriostatic, and features two way stretch. It is removable and easy to clean. The two-way stretch, reduces shear and friction.
Servicing
Service infra-red link - for system tracking and servicing.
Dimensions
AU-1000 Autoexcel Pump
Overall Length (mm) 336
Overall Height (mm) 210
Overall Width (mm) 100
Weight (kg) 3.5
Power 220-240V,50Hz
Power consumption 25 VA
Electrical Safety Standards IEC 601-1
BS 5724 part 1
VDE 0750:Teil 1)
AU-1020 Autoexcel Mattress (Inflated Dimensions)
Overall Length (mm) 1980
Height at edges (mm) 135
Overall Width (mm) 850
Weight Limit (kg) 203
Additional Comment
- Remote control unit can recall data on the systems performance.
- Information can be obtained on recent alarm history, running hours since last service and total operating hours.
- The dynamic system has over 280 air changes per day.
- Fits standard hospital bed or single bed.
- PUMP has an on / off switch and an automatically operated pressure control. A light and audible alarm indicates low pressure.
- PUMP / MATTRESS CONNECTORS have a "snap lock" mechanism to minimise accidental disconnection.
- The EMERGENCY RELEASE Cardio Pulmonary Resuscitation (CPR) is possible without having to remove the overlay from the bed. A pull ring release deflates the mattress in under 10 seconds to enable resuscitation procedures to be performed.
- The Autoexcel Mattress Overlay System is lightweight, portable and easy to use.
Warranty
- 2 years extended warranty on all parts and labour.
Maintenance
The overlay may be sponged down. Infection control guidelines are available. The pumps can be repaired. If an air chamber is punctured the whole air chamber can be removed and a replacement chamber press studded into place.
Place Of Manufacture
HUNTLEIGH HEALTHCARE (UK)
Supplier Information - click to supplier for contacts details
ARJOHUNTLEIGH,
NEWINGTON, NSW
Additional /images
No additional /images apply to this item
First
Prev
Next
Last